Dental chair installation
Dental chair

Today we are installing a dental chair. We provide a white glove installation services.


Installing a dental chair with water, electrical, and air hookup requires careful planning and execution to ensure the safety and functionality of the system. Here are the steps to follow for a successful installation:


1) Determine the location: The first step in installing a dental chair with water, electrical, and air hookup is to choose the location where the chair will be installed. This should be a space that is easily accessible to the necessary utilities, including water supply, compressed air supply, and electrical power. The flooring should also be suitable for heavy equipment.


2) Install the utilities: The next step is to install the necessary utilities. This may involve hiring a professional plumber, electrician, or HVAC technician, depending on the complexity of the installation. In general, the following steps will be required:


a) Plumbing: A water supply line must be installed to provide water to the dental unit. This line may need to be run from a nearby source or extended from an existing water line. The line should be connected to a shut-off valve and a water pressure regulator to ensure consistent water pressure. A drain line should also be installed to remove wastewater from the dental unit.


b) Electrical: A dedicated electrical circuit is required to power the dental unit. This may involve running new electrical wiring or upgrading the existing electrical system. The electrical panel should be equipped with a circuit breaker to prevent overloading the circuit.


c) Compressed air: A compressed air supply line is required to power the handpieces and other pneumatic components of the dental unit. The line should be connected to an air compressor and a pressure regulator to ensure consistent air pressure.

Install the dental chair base: Once the utilities are in place, the dental chair base should be installed securely to the floor using bolts and anchors. The base should be leveled to ensure stability and proper function of the hydraulic lift system.


4) Install the dental unit: The dental unit, which includes the handpieces, water supply, air compressor, and suction system, should be installed according to the manufacturer's instructions. This may involve assembling the unit and connecting components such as hoses, fittings, and valves. Ensure that all connections are made securely and that the system is functioning properly before proceeding.


5) Connect the utilities: The next step is to connect the utilities to the dental unit. This may involve connecting water lines, electrical wiring, and air hoses, depending on the system. Make sure that all connections are made securely and that there are no leaks or other issues.


6) Test the system: Before using the dental chair on patients, it is important to test the entire system, including the water supply, air supply, and electrical connections. Check for leaks or other issues and make sure that all components are functioning properly. Test the handpieces and suction system to ensure proper operation.


7) Train staff: Finally, it is important to train staff on how to use the new equipment properly and safely. Make sure that all staff members are familiar with the location of shut-off valves, emergency stop buttons, and other safety features. Review the manufacturer's instructions and ensure that all staff members understand how to use the various components of the dental unit.


Overall, installing a dental chair with water, electrical, and air hookup requires careful planning and execution to ensure that the system is safe and functional. It is recommended to hire a professional dental equipment installer or technician to ensure proper installation and functionality of the system.